Negligible Specimen Hemolysis Is Observed With Sarstedt S-Monovette Coagulation Tubes in Aspiration Mode.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 35849104.
- Also identified by DOI 10.1093/ajcp/aqac071.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We aimed to compare the levels of hemolysis in the blood collected using the vacuum and aspiration modes via Sarstedt S-Monovette coagulation tubes. Forty volunteers were included in the study. Blood samples were collected using two different modes in the S-Monovette citrate tube (Sarstedt AG). Prothrombin time, active partial thromboplastin time, fibrinogen, and D-dimer analyses were performed using the STA-Compact-Max 3 analyzer (Stago). The hemolysis levels of the samples were measured by both Stago's semiquantitative hemolysis index (H-index) module and the quantitative H-index measurement of the Roche cobas 6000 (Roche Diagnostics) analyzer. Roche's quantitative H-index values were statistically significantly lower in the aspiration mode. No clinically significant difference was observed between coagulation test results. Using the S-Monovette citrate tubes can reduce spurious hemolysis and improve patient safety.
Medical subject headings
- Hematologic Tests
- Hemolysis