The Relationship Between Different Bench Test Methodologies and Accuracy of Insulin Infusion Pumps: A Systematic Literature Review.
systematic_review · Level I
Where this comes from
- Record sourced from PubMed, PMID 35953739.
- Also identified by DOI 10.1007/s10439-022-03027-7.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Diabetes technology has rapidly evolved, and insulin infusion pumps (IIPs) have gained worldwide acceptance in diabetes care. The safety of medical equipment is highly discussed, imposing complex challenges in its use. The accuracy of IIPs can be determined through laboratory tests, generally following the IEC 60601-2-24 protocol. Studies have evaluated the accuracy and precision of IIPs, and there are discrepant results. So, we conducted a Systematic Literature Review to assess the methodologies used to evaluate the accuracy of IIPs, organizing the findings in a compiled perspective. The methodology was based on Kitchenham and Biolchini guidelines, and when possible it was carried out the Bayesian meta-analyses to compare the accuracy of IIPs. Most studies used the microgravimetric technique to evaluate the device accuracy, and some proposed adaptations for the standard protocol. The variation of results was recurrent, and the establishment of a protocol, especially to evaluate patch pumps, is necessary. The present study gives enough data to understand the scenario of the IIPs evaluation, as well as the different protocols that can be explored for its evaluation. This highlights the need for a reliable, practical, and low-cost methodology to assist the evaluation of IIPs.
Medical subject headings
- Diabetes Mellitus
- Insulin