Correlations between B-mode ultrasound image texture features and tissue temperatures in hyperthermia.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 36201496.
- Also identified by DOI 10.1371/journal.pone.0266446 and PMC identifier 9536603.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The noninvasive monitoring of mild hyperthermia or thermal ablation is important to guarantee therapeutic safety and efficacy. The potential of ultrasound B-mode image texture features in monitoring temperature or coagulation zones studied in this article. The experiments carried out on eighteen in vitro porcine liver samples heated from 20°C to 60°C in the water bath. The ultrasound radiofrequency signal at different temperature collected to reconstruct B-mode ultrasound images. The texture features based on gray level histogram (GLH), gray level co-occurrence matrix (GLCM), and gray level-gradient co-occurrence matrix (GGCM) extracted, respectively. Accordingly, we analyze the correlations between these texture features and temperature based on the experiment results. The results showed that five texture feature parameters closely related to temperature, including mean gray scale of GLH, homogeneity of GLCM, hybrid entropy, inverse difference moment, and correlation of GGCM. Some of these feature parameters have correlation coefficients larger than 0.9 within the temperature range of 20°C to 60°C. The above-mentioned five feature parameters expected to apply for noninvasive monitoring of MH or TA.
Medical subject headings
- Hyperthermia, Induced