Abdominopelvic CT Image Quality: Evaluation of Thin (0.5-mm) Slices Using Deep Learning Reconstruction.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 36259592.
- Also identified by DOI 10.2214/AJR.22.28319.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
<b>BACKGROUND.</b> Because thick-section images (typically 3-5 mm) have low image noise, radiologists typically use them to perform clinical interpretation, although they may additionally refer to thin-section images (typically 0.5-0.625 mm) for problem solving. Deep learning reconstruction (DLR) can yield thin-section images with low noise. <b>OBJECTIVE.</b> The purpose of this study is to compare abdominopelvic CT image quality between thin-section DLR images and thin- and thick-section hybrid iterative reconstruction (HIR) images. <b>METHODS.</b> This retrospective study included 50 patients (31 men and 19 women; median age, 64 years) who underwent abdominopelvic CT between June 15, 2020, and July 29, 2020. Images were reconstructed at 0.5-mm section using DLR and at 0.5-mm and 3.0-mm sections using HIR. Five radiologists independently performed pairwise comparisons (0.5-mm DLR and either 0.5-mm or 3.0-mm HIR) and recorded the preferred image for subjective image quality measures (scale, -2 to 2). The pooled scores of readers were compared with a score of 0 (denoting no preference). Image noise was quantified using the SD of ROIs on regions of homogeneous liver. <b>RESULTS.</b> For comparison of 0.5-mm DLR images and 0.5-mm HIR images, the median pooled score was 2 (indicating a definite preference for DLR) for noise and overall image quality and 1 (denoting a slight preference for DLR) for sharpness and natural appearance. For comparison of 0.5-mm DLR and 3.0-mm HIR, the median pooled score was 1 for the four previously mentioned measures. These assessments were all significantly different (<i>p</i> < .001) from 0. For artifacts, the median pooled score for both comparisons was 0, which was not significant for comparison with 3.0-mm HIR (<i>p</i> = .03) but was significant for comparison with 0.5-mm HIR (<i>p</i> < .001) due to imbalance in scores of 1 (<i>n</i> = 28) and -1 (slight preference for HIR, <i>n</i> = 1). Noise for 0.5-mm DLR was lower by mean differences of 12.8 HU compared with 0.5-mm HIR and 4.4 HU compared with 3.0-mm HIR (both <i>p</i> < .001). <b>CONCLUSION.</b> Thin-section DLR improves subjective image quality and reduces image noise compared with currently used thin- and thick-section HIR, without causing additional artifacts. <b>CLINICAL IMPACT.</b> Although further diagnostic performance studies are warranted, the findings suggest the possibility of replacing current use of both thin- and thick-section HIR with the use of thin-section DLR only during clinical interpretations.
Medical subject headings
- Deep Learning