The effect of trunnionosis on the risk of re-revision following femoral head exchange in hip arthroplasty.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 36357353.
- Also identified by DOI 10.1177/11207000221136204.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Fretting corrosion at modular junctions contributes to arthroplasty failure. Currently, no evidence-based guidelines are available regarding the acceptable level of trunnion corrosion that can occur <i>in vivo</i>. We aimed to examine the relationship between trunnion corrosion and risk of re-revision to assist surgeons with intraoperative decision making. Grading by 3 independent examiners of revised and re-revised head components was performed using a modified Goldberg corrosion scale. Samples were separated into low-grade (LG) and high-grade (HG) corrosion. Mechanical testing determined the relationship between corrosion severity and pull-off strength at the head-stem junction. 529 retrieved femoral heads were analysed. A positive association was detected between males and HG corrosion (OR 2.07; 95% CI, 1.45-2.94; <i>p</i> < 0.001). No difference between the survivorship of LG and HG heads was detected (<i>p</i>-value = 0.247). In the re-revised sample, the first implant had a time <i>in situ</i> that was on average 7.97 years longer (95% CI, 5.4-10.6) than that of the subsequent re-revised femoral head. Severe corrosion on the first head was associated with a 37.5 (95% CI, 4.00-1944) fold increase of HG on the subsequent head (<i>p</i> < 0.001). Femoral disassembly force had a positive correlation with stem taper corrosion grade (<i>p</i> = 0.001). A well-fixed stem with corrosion may remain <i>in situ</i>.
Medical subject headings
- Arthroplasty, Replacement, Hip
- Hip Prosthesis
Anatomy
- femur
- hip