Thickness-Sensing Sandwiched Plasmonic Biosensors Enable Label-Free Naked-Eye Antibody Quantification.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 36394551.
- Also identified by DOI 10.1021/acs.nanolett.2c03732 and PMC identifier 9805804.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Clinical serology assays for detecting the antibodies of the virus are time-consuming, are less sensitive/selective, or rely on sophisticated detection instruments. Here, we develop a sandwiched plasmonic biosensor (SPB) for supersensitive thickness-sensing via utilizing the distance-dependent electromagnetic coupling in sandwiched plasmonic nanostructures. SPBs quantitatively amplify the thickness changes on the nanoscale range (sensitivity: ∼2% nm<sup>-1</sup>) into macroscopically visible signals, thereby enabling the rapid, label-free, and naked-eye detection of targeted biomolecular species (via the thickness change caused by immunobinding events). As a proof of concept, this assay affords a broad dynamic range (7 orders of magnitude) and a low LOD (∼0.3 pM), allowing for the extremely accurate SARS-CoV-2 antibody quantification (sensitivity/specificity: 100%/∼99%, with a portable optical fiber device). This strategy is suitable for high-throughput multiplexed detection and smartphone-based sensing at the point-of-care, which can be expanded for various sensing applications beyond the fields of viral infections and vaccination.
Medical subject headings
- COVID-19
- Biosensing Techniques