Precision and accuracy of measuring finger motion with a depth camera: a cross-sectional study of healthy participants.
cross_sectional · Level IV
Where this comes from
- Record sourced from PubMed, PMID 36420794.
- Also identified by DOI 10.1177/17531934221138924.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The purpose of this cross-sectional study was to determine the precision and accuracy of the measurement of finger motion with a depth camera. Fifty-five healthy adult hands were included. Measurements were done with a depth camera and compared with traditional manual goniometer measurements. Repeated measuring showed that the overall repeatability and reproducibility of extension measured with the depth camera were within 3° and 4° and that of flexion were within 13° and 14°. Compared with traditional manual goniometry, biases of extension of all finger joints and flexion of metacarpophalangeal joints were less than 5°, and the average bias of flexion of proximal and distal interphalangeal joints was 29°. We conclude that the measurement of finger extension and flexion of the metacarpophalangeal joints with a depth camera was reliable, but improvement is required in the precision and accuracy of interphalangeal joint flexion.