Stent imaging on a clinical dual-source photon-counting detector CT system-impact of luminal attenuation and sharp kernels on lumen visibility.

Decker, Josua A; O'Doherty, Jim; Schoepf, U Joseph; Todoran, Thomas M; Aquino, Gilberto J; Brandt, Verena; Baruah, Dhiraj; Fink, Nicola et al. · Eur Radiol · 2023

basic_science · Level V

Where this comes from

Abstract

To assess the impact of scan modes and reconstruction kernels using a novel dual-source photon-counting detector CT (PCD-CT) on lumen visibility and sharpness of different stent sizes. A phantom containing six balloon-expandable stents (2.5 to 9 mm diameter) in silicone tubing was scanned on a PCD-CT with standard (0.6 mm and 0.4 mm thicknesses) and ultra-high-resolution (0.2 mm thickness) modes. With the use of increasing contrast medium concentrations, densities of 0, 200, 400, and 600 HU were achieved. Standard-resolution scans were reconstructed using increasing sharpness kernels, using both polyenergetic quantitative soft tissue "conventional" ((Qr40<sub>c</sub>(0.6 mm), Qr40<sub>c</sub>(0.4 mm), Qr72<sub>c</sub>(0.2 mm)) and vascular (Bv) virtual monoenergetic reconstructions (Bv44<sub>m</sub>(0.4 mm), Bv60<sub>m</sub>(0.4 mm)) at 70 keV. In-stent lumen visibility, sharpness (max. ΔHU of the stent measured in profile plots), and in-stent noise (standard deviation of HU) were measured. In-stent lumen visibility was highest for Qr72<sub>c</sub>(0.2 mm) (86.5 ± 2.8% to 88.3 ± 2.6%) and in Bv60<sub>m</sub>(0.4 mm) reconstructions (77.3 ± 2.9 to 82.7 ± 2.5%). Lumen visibility was lowest in the smallest stent (2.5 mm) ranging from 54.1% in Qr40<sub>c</sub>(0.6 mm) to 74.1% in Qr72<sub>c</sub>(0.2 mm) and highest in the largest stent (9 mm) ranging from 93.8% in Qr40c(0.6 mm) to 99.1% in the Qr72<sub>c</sub>(0.2 mm) series. Lumen visibility decreased by 2.1% for every 200-HU increase in lumen attenuation. Max. ΔHU between stents and stent lumen was highest in Qr72<sub>c</sub>(0.2 mm) (ΔHU 892 ± 504 to 1526 ± 517) and Bv60<sub>m</sub>(0.4 mm) series (ΔHU 480 ± 357 to 1030 ± 344). Improvement of lumen visibility and sharpness in UHR and Bv60<sub>m</sub>(0.4 mm) series was strongest in smaller stent sizes. UHR acquisition mode and sharp reconstruction kernels on a novel PCD-CT system significantly improve in-stent lumen visibility and sharpness-especially for smaller stent sizes. • In-stent lumen visibility and sharpness of stents significantly improve using sharp reconstruction kernels (Bv60) and ultra-high-resolution mode in photon-counting detector computed tomography. • The observed improvement of stent-lumen visibility was highest in smaller stent sizes.

Medical subject headings