Design and quantitative evaluation of 'Aerosol Bio-Containment Device (ABCD)' for reducing aerosol exposure during infectious aerosol-generating events.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 36608021.
- Also identified by DOI 10.1371/journal.pone.0272716 and PMC identifier 9821519.
- Licence recorded as CC BY.
- The licence permits redistribution, so the abstract is shown in full and the full text is available from the publisher.
Abstract
The Coronavirus Disease 2019 (COVID-19) pandemic renewed interest in infectious aerosols and reducing risk of airborne respiratory pathogen transmission, prompting development of devices to protect healthcare workers during airway procedures. However, there are no standard methods for assessing the efficacy of particle containment with these protective devices. We designed and built an aerosol bio-containment device (ABCD) to contain and remove aerosol via an external suction system and tested the aerosol containment of the device in an environmental chamber using a novel, quantitative assessment method. The ABCD exhibited a strong ability to control aerosol exposure in experimental and computational fluid dynamic (CFD) simulated scenarios with appropriate suction use and maintenance of device seals. Using a log-risk-reduction framework, we assessed device containment efficacy and showed that, when combined with other protective equipment, the ABCD can significantly reduce airborne clinical exposure. We propose this type of quantitative analysis serves as a basis for rating efficacy of aerosol protective enclosures.