Sensor selection and miniaturization limits for detection of interictal epileptiform discharges with wearable EEG.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 36630712.
- Also identified by DOI 10.1088/1741-2552/acb231.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
<i>Objective.</i>The goal of this paper is to investigate the limits of electroencephalography (EEG) sensor miniaturization in a set-up consisting of multiple galvanically isolated EEG units to record interictal epileptiform discharges (IEDs), referred to as 'spikes', in people with epilepsy.<i>Approach.</i>A dataset of high-density EEG recordings (257 channels) was used to emulate local EEG sensor units with short inter-electrode distances. A computationally efficient sensor selection and interictal spike detection algorithm was developed and used to assess the influence of the inter-electrode distance and the number of such EEG units on spike detection performance. Signal-to-noise ratio, correlation with a clinical-grade IEDs detector and Cohen's kappa coefficient of agreement were used to quantify performance. Bayesian statistics were used to confirm the statistical significance of the observed results.<i>Main results.</i>We found that EEG recording equipment should be specifically designed to measure the small signal power at short inter-electrode distance by providing an input referred noise<300 nV. We also found that an inter-electrode distance of minimum 5 cm between electrodes in a setup with a minimum of two EEG units is required to obtain near equivalent performance in interictal spike detection to standard EEG.<i>Significance.</i>These findings provide design guidelines for miniaturizing EEG systems for long term ambulatory monitoring of interictal spikes in epilepsy patients.
Medical subject headings
- Epilepsy
- Wearable Electronic Devices