Surface and mechanical properties of polyurethane central venous catheters after repeated contact with chemotherapy excipient solutions.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 36705442.
- Also identified by DOI 10.1002/jbm.b.35224.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
This article investigates the impact of the interactions between polyurethane central venous catheters and solutions containing excipients used in cisplatin and paclitaxel formulations. Changes to the properties of catheters and the leaching of catheter additives into the infused solutions were studied while these solutions were infused cyclically for several months. Chemotherapy treatment was mimicked in vitro in compliance with hospital practices. The treatment cycle was repeated 10 times, using solutions containing only the excipients. After 10 treatment cycles, no physical or chemical degradation of the catheter was observed. Mechanical performances were stable, but surface modifications occurred, causing the surface to become more hydrophobic. A loss in polyurethane antioxidant amount was observed in part due to a leaching phenomenon.
Medical subject headings
- Central Venous Catheters
- Catheterization, Central Venous