Evaluating a Novel, Portable, Self-Administrable Device ("Beacon") That Measures Critical Flicker Frequency as a Test for Hepatic Encephalopathy.

Vutien, Philip; Li, Richard; Karkar, Ravi; Munson, Sean A; Fogarty, James; Walter, Kara; Yacoub, Michael; Ioannou, George N · Am J Gastroenterol · 2023

cross_sectional · Level IV

Where this comes from

Abstract

We compared critical flicker frequency (CFF) thresholds obtained using a novel portable device "Beacon" with thresholds from the commercially available Lafayette Flicker Fusion System (Lafayette-FFS) in patients with cirrhosis. One hundred fifty-three participants with chronic liver disease underwent CFF testing using Beacon and Lafayette-FFS with a method-of-limits and/or forced-choice protocol. Beacon demonstrated excellent test-retest reliability (intraclass correlation 0.91-0.97) and good correlation with the Lafayette-FFS values (intraclass correlation 0.77-0.84). Forced-choice CFF were on average 4.1 Hz higher than method-of-limits descending CFFs. Beacon can be self-administered by patients with chronic liver disease and cirrhosis to measure CFF, a validated screening test for minimal hepatic encephalopathy.

Medical subject headings