Transcutaneous carbon dioxide monitoring during prolonged apnoea with high-flow nasal oxygen.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 36760034.
- Also identified by DOI 10.1111/aas.14216.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The duration of apnoeic oxygenation with high-flow nasal oxygen is limited by hypercapnia and acidosis and monitoring of arterial carbon dioxide level is therefore essential. We have performed a study in patients undergoing prolonged apnoeic oxygenation where we monitored the progressive hypercapnia with transcutaneous carbon dioxide. In this paper, we compared the transcutaneous carbon dioxide level with arterial carbon dioxide tension. This is a secondary publication based on data from a study exploring the limits of apnoeic oxygenation. We compared transcutaneous carbon dioxide monitoring with arterial carbon dioxide tension using Bland-Altman analyses in anaesthetised and paralysed patients undergoing prolonged apnoeic oxygenation until a predefined limit of pH 7.15 or PCO<sub>2</sub> of 12 kPa was reached. We included 35 patients with a median apnoea duration of 25 min. Mean pH was 7.14 and mean arterial carbon dioxide tension was 11.2 kPa at the termination of apnoeic oxygenation. Transcutaneous carbon dioxide monitoring initially slightly underestimated the arterial tension but at carbon dioxide levels above 10 kPa it overestimated the value. Bias ranged from -0.55 to 0.81 kPa with limits of agreement between -1.25 and 2.11 kPa. Transcutaneous carbon dioxide monitoring provided a clinically acceptable substitute for arterial blood gases but as hypercapnia developed to considerable levels, we observed overestimation at high carbon dioxide tensions in patients undergoing apnoeic oxygenation with high-flow nasal oxygen.
Medical subject headings
- Oxygen
- Apnea