Finite element analysis of a novel anatomical locking plate for scapular neck fracture.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 37004037.
- Also identified by DOI 10.1186/s13018-023-03743-3 and PMC identifier 10067195.
- Licence recorded as CC BY.
- The licence permits redistribution, so the abstract is shown in full and the full text is available from the publisher.
Abstract
Reconstruction plates (RPs) are commonly used in scapular neck fractures (SNFs): however, RPs have many defects. In this study, we evaluated a newly designed scapular neck anatomical locking compression plate (SNALCP). An SNF finite element model (Miller-type IIB) was constructed. Plates were subsequently implanted into the scapula and fixed with screws that were grouped according to the plate used: SNALCP (A) and RP (B). Finally, loads were applied to record and analyze performance. Under lateral, anteroposterior, and vertical compression loads, the maximum von Mises stresses on the scapula and implants of group A were smaller than those of group B. There were some differences in stress distribution between the two groups. SNALCP can effectively reduce the stress of the scapula and implant, making stress distribution more uniform and continuous, and has mechanical conduction advantages. Compared to RP, it provides improved stability and more reliable fixation.
Medical subject headings
- Fracture Fixation, Internal
- Shoulder Fractures
Anatomy
- scapula
- shoulder