Intra-rater reliability of the knee arthrometer GNRB® for measuring knee anterior laxity in healthy, active subjects.
cross_sectional · Level IV
Where this comes from
- Record sourced from PubMed, PMID 37089624.
- Also identified by DOI 10.1016/j.jor.2023.03.016 and PMC identifier 10113708.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Arthrometers are used to assess knee anterior laxity and to evaluate the integrity of the anterior cruciate ligament. Assessment of knee anterior laxity is crucial part of the clinical examination. The aim of this study was to investigate the intra-rater reliability of the GNRB® in healthy subjects. In the study participated 97 subjects and two measurements using a GNRB® arthrometer were performed. Males were tested two times one week a part. Females were tested two times within 24 h. Measurements were performed at the following forces 134 N, 134 N, 150 N, 200 N, and 200 N. To evaluate the reliability of GNRB® measurements an intraclass correlation coefficient was calculated. Intra-rater reliability of the GNRB® measurements is good for 134 N on the left knee (ICC = 0.848) and the right knee (ICC = 0.788) and for 200 N on the left knee (ICC = 0.805) and the right knee (ICC = 0.756). The GNRB® knee arthrometer has good intra-reliability for measurements at the 134 N and 200 N forces. Reliability can be increased with the standardize position of the subject, the stabilization of the patella and standardize measurement protocol.
Anatomy
- knee