The addition of junctions in modular stems implies a greater susceptibility to corrosion. The aim of this study is to compare serum chromium and cobalt levels after using a bimodular stem and its monoblock counterpart in primary total hip arthroplasty. Postoperative clinical scores were also compared. A prospective cohort study between 2012 and 2015 was designed. One arm of the cohort included patients with the cementless modular neck stem H-Max M® and the other with the cementless monoblock stem counterpart H-Max S®. No statistically significant difference was found in chromium value between groups (p = 0.621) at two years postoperative. Cobalt value was higher in the modular group (p = < 0.001). No statistically significant difference was found in clinical postoperative scores except for the Harris Hip Score, with better results at six months in modular group (p = 0.007). Higher serum cobalt level in the modular group has limited the use of modular stems in our daily practice. Advantages of modular stem were not found. II.