Case Study of Intrathecal Baclofen Pump Motor Shutdown Secondary to the Effect of the Magnetic Field Created by a Personal Digital Tablet and Magnetic Cover.

Fostier, Michele; Gustin, Thierry; Deltombe, Thierry · Am J Phys Med Rehabil · 2023

case_report · Level V

Where this comes from

Abstract

We present the case of a posttraumatic C6 AIS A tetraplegic patient with spasticity treated with an intrathecal baclofen pump, who noticed a transient increase in his spasticity each time he used a digital tablet (Ipad) protected by a magnetic shell placed on his abdomen. Telemetry confirmed transient motor shutdown responsible for withdrawal symptoms each time the tablet was used. Symptoms resolved after the removal of the protective shell. Effects of magnetic fields like magnetic resonance imaging are known to stall the pump rotor, which recover at the end of magnetic resonance imaging. Other sources of magnetic fields like laptops or new smartphones with magnet charging technology may also interfere with implanted devices. We therefore recommend patients to avoid close contact of magnetic devices with the intrathecal baclofen pump. More robust studies are warranted to assess the effect of the new magnetic technologies on the function of intrathecal pumps.