Concurrent Validity of the IOPI and Tongueometer Orofacial Strength Measurement Devices.

Curtis, James A; Mocchetti, Valentina; Rameau, Anaïs · Laryngoscope · 2023

cross_sectional · Level IV

Where this comes from

Abstract

This study examined the concurrent validity of two orofacial strength manometers: (1) the Iowa Oral Performance Instrument (IOPI) - the current, gold standard orofacial manometer; and (2) the Tongueometer - a newly-available, lower cost, orofacial manometer. This study compared IOPI and Tongueometer pressure readings across three experimental conditions. Experiment 1 compared full setup (manometer + tongue bulb) pressure readings between the IOPI and Tongueometer. Experiment 2 compared IOPI tongue bulb and Tongueometer tongue bulb pressure readings, while controlling for manometer. Experiment 3 compared IOPI manometer and Tongueometer manometer pressure readings, while controlling for tongue bulb. Pressures were applied manually within a laboratory setting. Lin's concordance correlation (ρ<sub>c</sub> ) was used to calculate level of agreement, with ρ<sub>c</sub> interpreted as 'poor' if <0.90, 'moderate' if 0.90 to <0.95, 'substantial' if 0.95 to <0.99, and 'excellent' if ≥0.99. 539 trials were analyzed. There was a median absolute difference of 2.4 kPa in pressure readings between the IOPI and Tongueometer full setups (manometer + tongue bulb). Correlations revealed substantial agreement between IOPI and Tongueometer full setups (experiment 1: n = 292; ρ<sub>c</sub>  = 0.986), tongue bulbs (experiment 2: n = 146; ρ<sub>c</sub>  = 0.987-0.992), and manometers (experiment 3: n = 101; ρ<sub>c</sub>  = 0.970). Differences in pressures were consistently observed between the Tongueometer and IOPI. Despite these differences, substantial agreement was present. These data suggest the Tongueometer may be a valid, lower cost alternative to the IOPI for objectively assessing orofacial strength in clinical practice. Level 2 Laryngoscope, 133:3123-3131, 2023.

Medical subject headings