Magnetic Resonancy Imaging Safety in Active Osseointegrated (Osia®) and Cochlear Implants: New Technology Creating Confusion.

Abdalla, Ibrahim; Choi, Janet S; Struyk, Griffin; Adams, Meredith E; Huang, Tina C · Laryngoscope · 2024

case_report · Level V

Where this comes from

Abstract

There is increased confusion regarding MRI-compatible CIs and BAHAs. This report describes two cases when patients underwent MRIs with non-MRI compatible devices. One patient with bilateral Cochlear Osias experienced dislocation of both internal magnets after 1.5 Tesla MRI. Both magnets were outside the silastic sheath, with the left magnet flipped. A second patient with a legacy CI experienced similar internal magnet dislocation and inversion after 3 Tesla MRI. This study describes internal magnet dislocation/inversion with the Cochlear Osia and a legacy CI after MRI. Our findings suggest the need for improved patient education and simplified radiology guidelines. Laryngoscope, 134:393-396, 2024.