Low complication rate and high implant survival at 2 years after Touch® trapeziometacarpal joint arthroplasty.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 37310049.
- Also identified by DOI 10.1177/17531934231179581.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We analysed complications, revision surgeries, and patient-reported and clinical outcomes 2 years after trapeziometacarpal joint implant arthroplasty using the Touch® prosthesis. Of 130 operated patients with trapeziometacarpal joint osteoarthritis, four had to be revised owing to implant dislocation, loosening or impingement, leading to an estimated 2-year survival rate of 96% (95% confidence interval: 90 to 99). Of 101 patients available for the 2-year follow-up, complications occurred in 17, with the most frequent being de Quervain stenosing vaginosis (<i>n</i> = 6) and trigger thumb (<i>n</i> = 5). Pain at rest decreased significantly from a median value of 5 (interquartile range [IQR]: 4 to 7) before surgery to 0 (IQR: 0 to 1) at 2 years. Key pinch strength increased significantly from 4.5 kg (IQR: 3.0 to 6.5) to 7.0 kg (IQR: 6.0 to 8.0). We recommend surgery with the Touch® prosthesis as the standard procedure for patients with isolated trapeziometacarpal joint osteoarthritis because of the high survival rate and promising outcomes at 2 years.<b>Level of evidence:</b> IV.
Medical subject headings
- Osteoarthritis
- Joint Prosthesis
- Carpometacarpal Joints
- Trapezium Bone
- Arthroplasty, Replacement
Anatomy
- hand
- wrist