Adverse Hemodynamic Consequences of Continuous Left Ventricular Mechanical Support: JACC Review Topic of the Week.
review · Level V
Where this comes from
- Record sourced from PubMed, PMID 37380306.
- Also identified by DOI 10.1016/j.jacc.2023.04.045.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Left ventricular assist devices (LVADs) provide lifesaving therapy for patients with advanced heart failure. The recognition of pump thrombosis, stroke, and nonsurgical bleeding as hemocompatibility-related adverse events (HRAEs) led to pump design improvements and reduced adverse event rates. However, continuous flow can predispose patients to right-sided heart failure (RHF) and aortic insufficiency (AI), especially as patients live longer with their device. Given the hemodynamic contributions to AI and RHF, these comorbidities can be classified as hemodynamic-related events (HDREs). Hemodynamic-driven events are time dependent and often manifest later than HRAEs. This review examines the emerging strategies to mitigate HDREs, with a focus on defining best practices for AI and RHF. As we head into the next generation of LVAD technology, it is important to differentiate HDREs from HRAEs so that we can continue to advance the field and improve the true durability of the pump-patient continuum.
Medical subject headings
- Aortic Valve Insufficiency
- Heart Failure
- Heart-Assist Devices