The minimum required overlap length for tendon transfer A biomechanical study on human tendons.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 37540707.
- Also identified by DOI 10.1371/journal.pone.0289650 and PMC identifier 10403071.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
In tendon transfer surgeries sufficient stability of the tenorrhaphy is essential. In addition to the choice of a suitable technique, adequate overlap of donor and recipient tendons must be ensured. The aim of this study was to investigate the tensile strength with regard to tendon overlap of a recently published tenorrhaphy, termed Woven-Fridén (WF) tenorrhaphy, which displayed higher tensile strength and lower bulk when compared to the established Pulvertaft technique. For this purpose, WF tenorrhaphies with 1.5 cm, 2 cm, and 3 cm tendon overlap were performed and subsequently tested for different biomechanical properties by tensile testing. Among others, the parameters of ultimate load and stiffness were collected. Native tendons served as controls. A formula was derived to quantify the relation between tendon overlap and ultimate load. We observed that sufficient tensile strength (mean ultimate load of 217 N) is already given with a 2 cm tendon overlap. In addition, with more than 3 cm overlap length only little additional tensile strength is to be expected as the calculated ultimate load of 4 cm overlap (397 N) is approaching the plateau of the maximal ultimate load of 435 N (native tendons).
Medical subject headings
- Tendon Transfer
- Plastic Surgery Procedures