Rapid Bacterial Detection and Gram-Identification Using Bacterially Activated, Macrophage-Membrane-Coated Nanowired-Si Surfaces in a Microfluidic Device.

Liu, Sidi; Wang, Huibo; Yu, Le; Ren, Yijin; Bouma, Hjalmar R; Liu, Jian; van der Mei, Henny C; Busscher, Henk J · Nano Lett · 2023

basic_science · Level V

Where this comes from

Abstract

Bacterially induced sepsis requires rapid bacterial detection and identification. Hours count for critically ill septic patients, while current culture-based detection requires at least 10 h up to several days. Here, we apply a microfluidic device equipped with a bacterially activated, macrophage-membrane-coating on nanowired-Si adsorbent surfaces for rapid, bacterial detection and Gram-identification in bacterially contaminated blood. Perfusion of suspensions of Gram-negative or Gram-positive bacteria through a microfluidic device equipped with membrane-coated adsorbent surfaces detected low (<10 CFU/mL) bacterial levels. Subsequent, <i>in situ</i> fluorescence-staining yielded Gram-identification for guiding antibiotic selection. In mixed <i>Escherichia coli</i> and <i>Staphylococcus aureus</i> suspensions, Gram-negative and Gram-positive bacteria were detected in the same ratios as those fixed in suspension. Results were validated with a 100% correct score by blinded evaluation (two observers) of 15 human blood samples, spiked with widely different bacterial strains or combinations of strains, demonstrating the potential of the platform for rapid (1.5 h in total) diagnosis of bacterial sepsis.