Rapid Bacterial Detection and Gram-Identification Using Bacterially Activated, Macrophage-Membrane-Coated Nanowired-Si Surfaces in a Microfluidic Device.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 37611221.
- Also identified by DOI 10.1021/acs.nanolett.3c02686 and PMC identifier 10510579.
- Licence recorded as CC BY.
- The licence permits redistribution, so the abstract is shown in full and the full text is available from the publisher.
Abstract
Bacterially induced sepsis requires rapid bacterial detection and identification. Hours count for critically ill septic patients, while current culture-based detection requires at least 10 h up to several days. Here, we apply a microfluidic device equipped with a bacterially activated, macrophage-membrane-coating on nanowired-Si adsorbent surfaces for rapid, bacterial detection and Gram-identification in bacterially contaminated blood. Perfusion of suspensions of Gram-negative or Gram-positive bacteria through a microfluidic device equipped with membrane-coated adsorbent surfaces detected low (<10 CFU/mL) bacterial levels. Subsequent, <i>in situ</i> fluorescence-staining yielded Gram-identification for guiding antibiotic selection. In mixed <i>Escherichia coli</i> and <i>Staphylococcus aureus</i> suspensions, Gram-negative and Gram-positive bacteria were detected in the same ratios as those fixed in suspension. Results were validated with a 100% correct score by blinded evaluation (two observers) of 15 human blood samples, spiked with widely different bacterial strains or combinations of strains, demonstrating the potential of the platform for rapid (1.5 h in total) diagnosis of bacterial sepsis.
Medical subject headings
- Bacteria
- Sepsis