A Laboratory-Developed Assay for Clade II Human Mpox Virus on the Panther Fusion Open Access System.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 37681583.
- Also identified by DOI 10.1093/infdis/jiad347.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
This paper describes the methods for developing and optimizing a laboratory-developed assay (LDA) for detecting clade II human mpox virus using the automated Panther Fusion platform and Open Access software. Various concentrations of reagents in a primer-probe mix were tested to optimize the LDA. The LDA was validated using 10 previously characterized positive and 10 negative human mpox samples, resulting in 95% accuracy and 100% precision. The LDA resulted in 100% specificity among previously tested HSV1-, HSV2-, and VZV-positive human samples. Several spiked media extensions were also validated and achieved 98% accuracy and 100% precision across all collection media types. The assay's limit of detection was calculated to be 1.475 copies/reaction, and the polymerase chain reaction efficiency resulted in 89.87% (slope, -3.5911; R2 = 0.9947). The methods described here can be applied to the rapid optimization and development of LDAs for many possible pathogens of public health importance.
Medical subject headings
- Monkeypox virus
- Viruses