Impact of a selective lens dose reduction protocol in 3D rotational angiography on radiation exposure to the eye lens during cerebral angiography: a randomized controlled trial.
rct · Level II
Where this comes from
- Record sourced from PubMed, PMID 37739793.
- Also identified by DOI 10.1136/jnis-2023-020730.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We aimed to investigate the radiation dose to the eye lens (lens dose) during cerebral angiography and to evaluate the effectiveness of the lens dose reduction protocol for 3-dimensional rotational angiography (3D-RA) in reducing overall lens dose exposure. We conducted a randomized, controlled clinical trial at a tertiary hospital with patients undergoing cerebral angiography. The lens dose reduction protocol in 3D-RA involved raising the table to position the patient's eye lens away from the rotation axis. The lens dose was estimated by measuring the entrance surface air kerma using a photoluminescent glass dosimeter. The lens doses of 3D-RA, overall examination, and image quality were analyzed and compared between the two groups. A total of 20 participants (mean age, 58±9.4 years; including 12 men [60%]) were enrolled and randomly assigned to either the conventional group or the dose reduction group. The median lens dose in 3D-RA was significantly lower in the dose reduction group compared with the conventional group (1.1 mGy vs 4.5 mGy, p<0.001). The total dose was significantly lower in the dose reduction group (median of 7.5 mGy vs 10.2 mGy, p=0.003). In the conventional group, 3D-RA accounted for 46% of the total lens dose, while in the dose reduction group, its proportion decreased to 16%. No significant differences were observed in the image quality between the groups. The lens dose reduction protocol resulted in a significant reduction in the lens dose of the 3D-RA as well as entire cerebral angiography, while maintaining the image quality.
Medical subject headings
- Cerebral Angiography
- Radiation Dosage
- Imaging, Three-Dimensional
- Lens, Crystalline
- Radiation Exposure