A microprocessor stance and swing control orthosis improves balance, risk of falling, mobility, function, and quality of life of individuals dependent on a knee-ankle-foot orthosis for ambulation.
rct · Level II
Where this comes from
- Record sourced from PubMed, PMID 37752724.
- Also identified by DOI 10.1080/09638288.2023.2258342.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The C-Brace microprocessor stance and swing control orthosis was designed to overcome safety and functional limitations of traditional knee-ankle-foot orthoses (KAFOs) for individuals with lower limb paresis. However, a systematic comparison to established KAFO types has not been performed in a bigger sample. International multicenter, randomized, controlled, cross-over clinical trial. Legacy KAFO users at risk of falling were randomized to KAFO/C-Brace or C-Brace/KAFO use for three months with each orthosis. Primary outcome was balance assessed with the Berg Balance Scale (BBS). Secondary outcomes were falls, mobility, function, and quality of life. Intention-to-treat analysis with 102 participants. With the C-Brace, the BBS improved by 3.3 ± 6.3 points (<i>p</i> < 0.0001). Significantly fewer participants presented BBS scores <40 indicative of increased fall risk (16 vs. 36, <i>p</i> = 0.018). Mean falls reduced from 4.0 ± 16.8 to 1.1 ± 3.3 (<i>p</i> = 0.002). Outcomes for function, mobility, and quality of life showed significant improvements with the C-Brace. The improvements in fall risk and mobility can be attributed to the stumble recovery and controlled knee flexion during weight bearing of the C-Brace and have a positive impact on the quality of life of users. The C-Brace represents an option for KAFO users with increased fall risk and reduced mobility.
Medical subject headings
- Quality of Life
- Accidental Falls
- Postural Balance
- Cross-Over Studies
- Foot Orthoses