Biomechanical Performance of Total Wrist Arthrodesis Plates With and Without Arthrodesis of the Carpometacarpal Joint.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 37804161.
- Also identified by DOI 10.1177/15589447231198263 and PMC identifier 11833838.
- Licence recorded as CC BY.
- The licence permits redistribution, so the abstract is shown in full and the full text is available from the publisher.
Abstract
It is unknown whether total wrist arthrodesis (TWA) should be performed with or without arthrodesis of the carpometacarpal joint (CMCJ). The aim of this study is to compare CMCJ-spanning TWA plates using 3D printed wrist arthrodesis model with and without arthrodesis of the CMCJ. Total wrist arthrodesis plates mounted to 3D printed models were tested under a 4-N bending load at 4 Hz for 50 000 cycles, increased by 15% every 10 000 cycles until failure. Plates with arthrodesis CMCJ were stiffer and failed at a significantly greater load and number of cycles than plates mounted to models without CMCJ arthrodesis. The Synthes stainless steel locking TWA plate performed better than the Trimed plate applied to the model without CMCJ arthrodesis and the Acumed plate applied to the model with CMCJ arthrodesis. Based on these findings, we recommend arthrodesis of the CMCJ in TWA. Incorporation of the CMCJ in TWA may protect against plate failure. If arthrodesis of the CMCJ is not performed, plate removal should be considered before breakage occurs. IV.
Medical subject headings
- Arthrodesis
- Bone Plates
- Carpometacarpal Joints
- Wrist Joint
Anatomy
- wrist
- hand