Superior metal artifact reduction of tin-filtered low-dose CT in imaging of lumbar spinal instrumentation compared to conventional computed tomography.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 37804455.
- Also identified by DOI 10.1007/s00256-023-04467-5 and PMC identifier 10858831.
- Licence recorded as CC BY.
- The licence permits redistribution, so the abstract is shown in full and the full text is available from the publisher.
Abstract
To compare the image quality of low-dose CT (LD-CT) with tin filtration of the lumbar spine after metal implants to standard clinical CT, and to evaluate the potential for metal artifact and dose reduction. CT protocols were optimized in a cadaver torso. Seventy-four prospectively included patients with metallic lumbar implants were scanned with both standard CT (120 kV) and tin-filtered LD-CT (Sn140kV). CT dose parameters and qualitative measures (1 = worst,4 = best) were compared. Quantitative measures included noise, signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), and the width and attenuation of the most prominent hypodense metal artifact. Standard CT and LD-CT were assessed for imaging findings. Tin-filtered LD-CT was performed with 60% dose saving compared to standard CT (median effective dose 3.22 mSv (quartile 1-3: 2.73-3.49 mSv) versus 8.02 mSv (6.42-9.27 mSv; p < .001). Image quality of CT and tin-filtered low-dose CT was good with excellent depiction of anatomy, while image noise was lower for CT and artifacts were weaker for tin-filtered LD-CT. Quantitative measures also revealed increased noise for tin-filtered low-dose CT (41.5HU), lower SNR (2) and CNR (0.6) compared to CT (32HU,3.55,1.03, respectively) (all p < .001). However, tin-filtered LD-CT performed superior regarding the width and attenuation of hypodense metal artifacts (2.9 mm and -767.5HU for LD-CT vs. 4.1 mm and -937HU for CT; all p < .001). No difference between methods was observed in detection of imaging findings. Tin-filtered LD-CT with 60% dose saving performs comparable to standard CT in detection of pathology and surgery related complications after lumbar spinal instrumentation, and shows superior metal artifact reduction.
Medical subject headings
- Tomography, X-Ray Computed
- Tin
Anatomy
- lumbar spine