Resonant scanning is critical to high speed and in vivo imaging in many applications of laser scanning microscopy. However, resonant scanning suffers from well-known image artifacts due to scanner jitter, limiting adoption of high-speed imaging technologies. Here, we introduce a real-time, inexpensive and all electrical method to suppress jitter more than an order of magnitude below the diffraction limit that can be applied to most existing microscope systems with no software changes. By phase-locking imaging to the resonant scanner period, we demonstrate an 86% reduction in pixel jitter, a 15% improvement in point spread function with resonant scanning and show that this approach enables two widely used models of resonant scanners to achieve comparable accuracy to galvanometer scanners running two orders of magnitude slower. Finally, we demonstrate the versatility of this method by retrofitting a commercial two photon microscope and show that this approach enables significant quantitative and qualitative improvements in biological imaging.
Medical subject headings
- Image Processing, Computer-Assisted