Fabrication of a monorail-type diagnostic catheter for rescue retrieval technique of a distal embolic protection device.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 38285124.
- Also identified by DOI 10.1007/s00701-024-05951-0.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
In this technical report, we describe a challenging case concerning the retrieval of a distal embolic protection device (DEPD) post-carotid artery stenting. We propose a novel rescue retrieval technique for DEPD, employing a fabricated monorail-type HN5 diagnostic catheter. When integrated with existing strategies, this approach may optimize and streamline the process of DEPD removal.
Medical subject headings
- Carotid Stenosis
- Embolic Protection Devices