An EM-Tracked Approach for Calibrating the 3D Pose of Flexible Endoscopes.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 38402316.
- Also identified by DOI 10.1007/s10439-024-03469-1.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Flexible endoscopes are ideal instruments for visualizing and diagnosing the inner surfaces of organs via a minimally invasive incision. Calibrating a flexible endoscope is a troublesome yet inevitable process in image-based tools tracking. Aiming to simplify the calibration process, we propose an electromagnetic (EM)-tracked calibration approach that does not require any predefined poses of the EM sensor. A three-stage calibration protocol was presented in an extensor. First, the orientation of the endoscope tube was derived by conducting a circular rotation of the endoscope around its axis utilizing a pair of tightly bearing stands. Second, the 3D position of the endoscope tip was acquired by having the tip come into contact with a flat plane. Third, the pose model of the bending section was derived and transformed into the local coordinate system of the EM sensor attached to the endoscope handle. To assess the accuracy of the proposed calibration approach, two experiments were designed and performed. Experimental results indicate accuracies of 0.09 ± 0.06 deg and 0.03 ± 0.19 deg in the estimation of the endoscope tube orientation and 0.52 ± 0.29, 0.33 ± 0.11, and 0.29 ± 0.17 mm in the x, y, and z estimations of the endoscope tip position, respectively. The proposed approach is accurate and easy to operate, does not require the employment of custom calibration markers, and can be used not only in surgical training systems but also in the endoscopic-based tools tracking.
Medical subject headings
- Endoscopes
- Endoscopy