Reliability of two devices for shoulder strength assessment: Wii Fit Balance Board and hand-held dynamometer.
cross_sectional · Level IV
Where this comes from
- Record sourced from PubMed, PMID 38425742.
- Also identified by DOI 10.1177/17585732221145558 and PMC identifier 10901178.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The aim of this study was to compare the reliability and agreement between two devices - Wii Fit Balance Board (WBB) versus Hand-Held Dynamometer (HHD) to measure isometric strength during the athletic shoulder (ASH) test in healthy amateur rugby players. Fifteen males (23.73 ± 2.8 years) completed two testing sessions. Maximal isometric contractions using the dominant arm (D) and non-dominant arm (ND) against a WBB and HHD were assessed at three angles of abduction (180°, 'I'; 135°, 'Y' and 90°, 'T'), in a prone lying position. The results indicate a very large correlation between the HHD and the WBB. WBB provides acceptable reliability at I-Test D (CV = 9.97%, ICC = 0.88) and HHD in the I-Test D (CV = 8.90%, ICC = 0.94), I-Test ND (CV = 8.60%, ICC = 0.95) in peak strength values. The HHD is most reliable in D ASH I-Y-T (CV = 10.94%) and WBB (CV = 11.05%). In the ND ASH I-Y-T test, the HHD is the most reliable (CV = 12.5%) compared to the WBB (CV = 14.43%). These results suggest that WBB is a reliable device to assess strength in the ASH test with a very large correlation with the HHD. WBB and HHD are two affordable devices to assess isometric shoulder strength.