Melanin bias in pulse oximetry explained by light source spectral bandwidth.

Bierman, Andrew; Benner, Kevin; Rea, Mark S · Br J Anaesth · 2024

basic_science · Level V

Where this comes from

Abstract

Pulse oximetry uses noninvasive optical measurements of light transmission from each of two sources through vascularised living tissue over the cardiac cycle (SpO<sub>2</sub>). From those measurements, the relative amount of oxygenated haemoglobin (SaO<sub>2</sub>) in circulating blood can be deduced. Recent reports have shown that, compared with SaO<sub>2</sub> measurements from blood samples, SpO<sub>2</sub> measurements are biased erroneously high for patients with dark skin. We developed a new method, spectrally resolved photoplethysmography (srPPG), to examine how spectral bandwidth affects the transmission of polychromatic light through the fingertip across the cardiac cycle. We measured and recorded the spectral transmission through the fingertip as the O<sub>2</sub> concentration in inspired air was reduced. We applied digital spectral filters of two different bandwidths, narrow or broad, to the same srPPG recordings to determine whether SpO<sub>2</sub> readings systematically varied for the two bandwidths. The srPPG method also allowed us to measure the fractional amount of melanin in the optical path. The effect of melanin content on the ratio of SpO<sub>2</sub> readings for narrow and broad spectral bandwidths was analysed. We hypothesised, based upon the Beer-Lambert law, and then showed experimentally, that the light emission spectra of light-emitting diode light sources, as used in commercial pulse oximeters, result in erroneously high SpO<sub>2</sub> measurements for patients having greater melanin concentrations in their skin than those of the subject pool used for instrument calibration. To eliminate melanin bias, pulse oximeters should use much narrower spectral bandwidths than those used in current models.

Medical subject headings