Investigation of the short-term in vivo performance of metal-on-carbon fibre reinforced poly ether ether ketone Motec wrists: an explant analysis.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 38780046.
- Also identified by DOI 10.1177/17531934241249919 and PMC identifier 11699702.
- Licence recorded as CC BY.
- The licence permits redistribution, so the abstract is shown in full and the full text is available from the publisher.
Abstract
Total wrist arthroplasty (TWA) aims to restore pain-free motion to diseased joints. One such TWA, the Motec, has demonstrated good results with acceptable complication rates. It has recently been suggested that the metal-on-carbon fibre reinforced poly ether ether ketone (Mo-CFR-PEEK) version of the Motec TWA be implanted instead of the metal-on-metal version. An explant analysis was undertaken on seven Motec Mo-CFR-PEEK TWAs, revised for a variety of reasons, after a mean time of 2 years in vivo. Compared to a new Motec implant, five of the explanted metal heads and three of the CFR-PEEK cups became smoother in vivo, suggesting self-polishing and negative skewness, indicating some material loss in vivo. Two explanted cups showed indentation marks on their rims and one of these was from component impingement with embedded metallic debris. In the short-term, the articulating surfaces of Motec Mo-CFR-PEEK TWAs did not show major damage.<b>Level of evidence:</b> IV.