Interreader and Intrareader Reproducibility of <sup>18</sup>F-Flotufolastat Image Interpretation in Patients with Newly Diagnosed or Recurrent Prostate Cancer: Data from Two Phase 3 Prospective Multicenter Studies.

Kuo, Phillip H; Esposito, Giuseppe; Ulaner, Gary A; Yoo, Don; Zukotynski, Katherine; Ravizzini, Gregory C; Penny, Ross; Miller, Matthew P et al. · J Nucl Med · 2024

prospective_cohort · Level II

Where this comes from

Abstract

Interreader and intrareader reproducibility of <sup>18</sup>F-flotufolastat PET/CT scans in newly diagnosed and recurrent prostate cancer patients was assessed from masked image evaluations from two phase 3 studies. <b>Methods:</b> <sup>18</sup>F-flotufolastat PET/CT images of newly diagnosed (<i>n</i> = 352) or recurrent (<i>n</i> = 389) patients were evaluated by 3 masked readers. Cohen κ was used to assess pairwise patient- and region-level interreader agreement. Agreement among all readers was assessed using Fleiss κ. Intrareader agreement between the first and repeat read (20% of images, ≥4 wk later) was assessed using Cohen κ. <b>Results:</b> Pairwise interreader agreement was 95% or better (newly diagnosed) and 75% or better (recurrent). The κ coefficients were impacted by the high-agreement-low-κ paradox: Cohen κ ranged from not estimable to 0.55, whereas Fleiss κ was 0.50 (newly diagnosed) and 0.41 (recurrent). Agreement was highest in the prostate of newly diagnosed patients (≥95%) and in the pelvic lymph nodes in recurrent patients (≥87%). Intrareader agreement was 86% or better across both populations. <b>Conclusion:</b> <sup>18</sup>F-flotufolastat PET/CT images can be reliably interpreted, with a high degree of inter- and intrareader agreement.