Retrieval Analysis of Titanium Nitride Coatings for Orthopaedic Implants.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 39019413.
- Also identified by DOI 10.1016/j.arth.2024.07.001.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The first generation of titanium nitride (TiN) coatings for orthopaedic implants was clinically introduced in the 1990s because of their promising biocompatibility, wear resistance, and corrosion resistance. This study evaluated the in vivo performance of early TiN-coated knee and hip implants, focusing on the bearing surfaces and mechanisms of in vivo damage. There were 13 TiN-coated implants (5 knee and 8 hip) retrieved from 8 patients as part of a multi-institutional implant retrieval program. The average implantation time was 4.25 years for knees and 17.5 years for hips. Implant revisions occurred for various reasons, including polyethylene wear, loosening, pain, infection, and instability. Components were examined using a semiquantitative scoring method, and surface roughness measurements were performed using white-light interferometry. Surface morphology, chemistry, and particle characterization were also assessed by scanning electron microscopy. For hips, mild corrosion was found on femoral head tapers, along with severe scratching on certain femoral heads. Knee implants exhibited low burnishing and scratching for both mechanisms. Roughness measurements (S<sub>a</sub>) were 37.3 nm (interquartile range = 22.0 to 62.4) for hips and 85.3 nm (interquartile range = 66.3 to 110) for knees. The observed scratch depth in both hip and knee implants due to third-body particles ranged from 0.3 to 1.3 μm. The coating coverage remained intact in the majority of the implants, with 2 cases of small, localized cohesive chipping and substrate exposure. The results of this study confirm the potential in vivo durability of early TiN coatings and will be useful in benchmarking wear tests for modern TiN-coated orthopaedic implants.
Medical subject headings
- Humans
- Titanium
- Hip Prosthesis
- Knee Prosthesis
- Coated Materials, Biocompatible
- Coated Materials, Biocompatible/chemistry
- Aged
- Female
- Middle Aged
- Surface Properties
- Male
- Prosthesis Failure
- Device Removal
- Arthroplasty, Replacement, Hip
- Arthroplasty, Replacement, Hip/instrumentation
- Arthroplasty, Replacement, Knee
- Arthroplasty, Replacement, Knee/instrumentation
- Prosthesis Design
- Corrosion
- Reoperation
- Microscopy, Electron, Scanning
Anatomy
- hip
- knee