TEG® 6s coagulation testing with a novel heparin neutralization cartridge: Technical validation and determination of normal reference ranges.
Where this comes from
- Record sourced from PubMed, PMID 39073983.
- Also identified by DOI 10.1093/ajcp/aqae088 and PMC identifier 11775113.
- Licence recorded as CC BY-NC.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We sought to establish normal reference ranges (NRRs) for a novel TEG 6s cartridge (TEG 6s Citrated: K, KH, RTH, FFH [Global Hemostasis]) (Haemonetics Corporation, Boston, MA, US). Healthy volunteers (≥18 years of age) included in this single-arm study provided single samples of whole blood. Primary end points included TEG parameters in the citrated kaolin (CK), CK with heparinase (CKH), RapidTEG with heparinase (CRTH), and functional fibrinogen with heparinase (CFFH) assays. Evaluable data were contributed by 164 volunteers (48.8% female; 62% White/Caucasian). The following NRRs were established: CK maximum amplitude (MA), 51.0 to 67.6 mm; CKH-MA, 51.8 to 67.9 mm; CRTH-MA, 53.0 to 68.9 mm; CFFH-MA, 15.3 to 34.4 mm; CK reaction time, 5.0 to 9.1 minutes; CKH reaction time, 4.9 to 9.4 minutes; CKH lysis 30 minutes after MA, 0% to 3.2%. Duplicate measurements demonstrated high reproducibility. CFFH-MA correlated with Clauss fibrinogen concentration (Pearson correlation coefficient, 0.74). Laboratory-based studies demonstrated maintenance of the relationship between CFFH-MA and fibrinogen up to 1344 mg/dL (hyperfibrinogenemic samples) and acceptability of heparin neutralization up to concentrations of low molecular weight and unfractionated heparin of 1.3 IU/mL and 5 IU/mL, respectively. This study established NRRs for the Global Hemostasis cartridge and serves as a proof of concept for the validity of results obtained using this cartridge.
Medical subject headings
- Thrombelastography
- Heparin
- Blood Coagulation