Flexible/Regenerative Nanosensor with Automatic Sweat Collection for Cytokine Storm Biomarker Detection.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 39099110.
- Also identified by DOI 10.1021/acsnano.4c04456.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The real-time monitoring of low-concentration cytokines such as TNF-α in sweat can aid clinical physicians in assessing the severity of inflammation. The challenges associated with the collection and the presence of impurities can significantly impede the detection of proteins in sweat. This issue is addressed by incorporating a nanosphere array designed for automatic sweat transportation, coupled with a reusable sensor that employs a Nafion/aptamer-modified MoS<sub>2</sub> field-effect transistor. The nanosphere array with stepwise wettability enables automatic collection of sweat and blocks impurities from contaminating the detection zone. This device enables direct detection of TNF-α proteins in undiluted sweat, within a detection range of 10 fM to 1 nM. The use of an ultrathin, ultraflexible substrate ensures stable electrical performance, even after up to 30 extreme deformations. The findings indicate that in clinical scenarios, this device could potentially provide real-time evaluation and management of patients' immune status via sweat testing.
Medical subject headings
- Sweat
- Biomarkers
- Biosensing Techniques