An Accurate Non-Contact Photoplethysmography via Active Cancellation of Reflective Interference.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 39146172.
- Also identified by DOI 10.1109/JBHI.2024.3443988.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Imaging Photoplethysmography (IPPG) is an emerging and efficient optical method for non-contact measurement of pulse waves using an image sensor. While the contactless way brings convenience, the inevitable distance between the sensor and the subject results in massive specular reflection interference on the skin surface, which leads to a low Signal to Interference plus Noise Ratio (SINR) of IPPG. To ease this challenge, this work proposes a novel modulation illumination approach to measure the accurate arterial pulse wave via surface reflection interference isolation from IPPG. Based on the proposed skin reflection model, a specific modulation illumination is designed to separate the surface reflections and obtain the subcutaneous diffuse reflections containing the pulse wave information. Compared with the results under ambient illumination and constant supplemental illumination, the SINR of the proposed method is improved by 4.56 and 3.74 dB, respectively.
Medical subject headings
- Photoplethysmography
- Signal Processing, Computer-Assisted