Biological tissue for transcatheter aortic valve: The effect of crimping on fatigue strength.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 39276437.
- Also identified by DOI 10.1016/j.jmbbm.2024.106741.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Transcatheter aortic valve replacement (TAVR) has become today the most attractive procedure to relieve patients from aortic valve disease. However, the procedure requires crimping biological tissue within a metallic stent for low diameter catheter insertion purpose. This step induces specific stress in the leaflets especially when the crimping diameter is small. One concern about crimping is the potential degradations undergone by the biological tissue, which may limit the durability of the valve once implanted. The purpose of the present work is to investigate the mechanical damage undergone by bovine pericardium tissue during compression and analyze how this degradation evolves with time under fatigue testing conditions. Pericardium 500 μm thick pericardium ribbons (5 mm large, 70 mm long) were crimped down to 12 Fr for 30 and 50 min within a metallic stent to replicate the heart valve crimping configuration. After crimping, samples underwent cyclic fatigue flexure and pressure loading over 0.5 Mio cycles. Samples were characterized for mechanical performances before crimping, after crimping and after fatigue testing in order to assess potential changes in the mechanical properties of the tissue after each step. Results bring out that the ultimate tensile strength is not modified through the process. However an increase in the modulus shows that the crimping step tends to stiffen the pericardium. This may have an influence on the lifetime of the implant.
Medical subject headings
- Materials Testing
- Transcatheter Aortic Valve Replacement
- Pericardium