Biomechanical analysis of cadaveric wrists before and after MOTEC wrist arthroplasty using a hexapod robot.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 39283016.
- Also identified by DOI 10.1177/17531934241274135.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
This study compares wrist motion, biomechanical behaviour and radiographic parameters before and after total wrist arthroplasty using a fourth-generation spherical articulation prosthesis. A total of 10 cadaveric specimens were assessed using a hexapod Stewart platform robot. After arthroplasty, there were significant increases in both stiffness and phase angle of wrist motion across all planes of motion assessed. In three specimens, a sudden increase in moment was observed on load/displacement curves. Radiographically, carpal height increased by 14%, and the centre of rotation was displaced 11.1 mm proximally, 4.6 mm dorsally and 3.9 mm radially. This stretched the musculotendinous units, tightening the joint, while increasing the moment arm of the wrist flexors and decreasing the moment arm of the extensors, potentially important in the development of postoperative flexion contractures. Possible alterations in technique and/or implant design are considered to assist surgeons in achieving optimal clinical and survivorship outcomes.
Medical subject headings
- Wrist Joint
- Joint Prosthesis
- Arthroplasty, Replacement
- Robotic Surgical Procedures
Anatomy
- wrist