Diagnostic accuracy of hemodynamic assessment of intracranial atherosclerotic stenosis from a single angiographic view: a validation study.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 39304194.
- Also identified by DOI 10.1136/jnis-2024-022114.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The aim of this study is to assess the feasibility of identifying the hemodynamic status of intracranial atherosclerotic stenosis (ICAS) using angio-based fractional flow (FF) calculated from a single angiographic view, with wire-based FF as the reference standard. The study retrospectively recruited 100 ICAS patients who underwent pressure wire measurement and digital subtraction angiography. The AccuICAD software was used to calculate angio-based FF, with the wire-measured value serving as the reference standard for evaluating the accuracy, consistency, and diagnostic performance of angio-based FF. The mean±SD value of wire-based FF was 0.77±0.18, while the mean value of angio-based FF was 0.77±0.19. A good correlation between angio-based FF and wire-based FF was evident (r=0.90, P<0.001), with good agreement (mean difference 0.00±0.08). The diagnostic accuracy of angio-based FF and percent diameter stenosis (DS%) were 93.23% versus 72.18%, 91.73% versus 72.93%, and 89.47% versus 78.95% for predicted wire-based FF thresholds of 0.70, 0.75, and 0.80, respectively. The area under the curve (AUC) values for angio-based FF and DS% were 0.975 versus 0.822, 0.970 versus 0.814, and 0.943 versus 0.826 at the respective thresholds, respectively. The FF calculated from a single angiographic view can be considered an effective tool for functional assessment of cerebral arterial stenosis.
Medical subject headings
- Intracranial Arteriosclerosis
- Angiography, Digital Subtraction
- Cerebral Angiography
- Hemodynamics
- Cerebrovascular Circulation