An acoustic bellows-type round window transducer for middle-ear implants.

Shin, Dong Ho; Lee, Kyu-Yup; Seong, Ki Woong · Clin Biomech (Bristol) · 2024

biomechanical · Level V

Where this comes from

Abstract

This study describes the development of output devices for round window middle-ear. To overcome the problems of output devices that apply sound pressure directly to the round window, an acoustic bellows-type round window transducer was implemented by combining a small bellows, acoustic tube, and balanced armature driver. The output characteristics of the proposed acoustic bellows-type round window transducer were confirmed through bench tests and distortion measurements. To compare the vibration transmission characteristics of the proposed transducer with those of sound pressure stimulation devices, an experiment was performed using four human temporal bones. The average output magnitude of the acoustic bellows-type round window transducer was equivalent to sound pressure levels of 92, 96, and 108 dB for frequency ranges of <1, 1-2, and > 2 kHz, respectively. The results showed that the proposed transducer delivered vibration consistently without reducing the sound pressure level due to leakage, unlike the sound pressure stimulation device. Therefore, the acoustic bellows-type round window transducer is a more stable and suitable output device for round window middle-ear implants than a sound pressure stimulation device. It is expected to overcome the limitations of sound pressure stimulation devices and to contribute to new technical solutions in the field of round window middle-ear implants development.

Medical subject headings