Label-free high-throughput impedance-activated cell sorting.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 39315634.
- Also identified by DOI 10.1039/d4lc00487f.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Cell sorting holds broad applications in fields such as early cancer diagnosis, cell differentiation studies, drug screening, and single-cell sequencing. However, achieving high-throughput and high-purity in label-free single-cell sorting is challenging. To overcome this issue, we propose a label-free, high-throughput, and high-accuracy impedance-activated cell sorting system based on impedance detection and dual membrane pumps. Leveraging the low-latency characteristics of FPGA, the system facilitates real-time dual-frequency single-cell impedance detection with high-throughput (5 × 10<sup>4</sup> cells per s) for HeLa, MDA-MB-231, and Jurkat cells. Furthermore, the system accomplishes low-latency (less than 0.3 ms), label-free, high-throughput (1000 particles per s) and high-accuracy (almost 99%) single-particle sorting using FPGA-based high-precision sort-timing prediction. In experiments with Jurkat and MDA-MB-231 cells, the system achieved a throughput of up to 1000 cells per s, maintaining a pre-sorting purity of 28.57% and increasing post-sorting purity to 97.09%. These findings indicate that our system holds significant potential for applications in label-free, high-throughput cell sorting.
Medical subject headings
- Electric Impedance