Bending the rules: A novel approach to laryngeal surgery in a body donor study.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 39327800.
- Also identified by DOI 10.1002/hed.27939 and PMC identifier 11717936.
- Licence recorded as CC BY.
- The licence permits redistribution, so the abstract is shown in full and the full text is available from the publisher.
Abstract
Transoral laser microsurgery, the standard surgical approach for early-stage laryngeal cancer, necessitates an unobstructed line of sight to the operating field. However, achieving adequate laryngeal exposure can be challenging, potentially compromising treatment outcomes. We developed a 3D-printed curved laryngoscope (sMAC), designed to match the upper airway anatomy. In a user study (n = 15) with a human body donor we compared the sMAC system to conventional microlaryngoscopy regarding laryngeal exposure and accessibility in a difficult exposure scenario. All 15 participants achieved complete glottic exposure and successfully manipulated laryngeal landmarks using the sMAC system. Only four participants achieved partial exposure using microlaryngoscopy. Positioning of the sMAC system was significantly faster (p = 0.023). A vocal cord resection was conducted successfully (n = 2) using the sMAC system. The sMAC system effectively addresses challenges associated with transoral laryngeal surgery. Ongoing development aims to overcome current limitations of the system and prepare first clinical trials.
Medical subject headings
- Laryngoscopy