Placing the tibial component of an ankle prosthesis: results of a CT-based simulation in healthy adults.
cross_sectional · Level IV
Where this comes from
- Record sourced from PubMed, PMID 39331136.
- Also identified by DOI 10.1007/s00590-024-04112-4 and PMC identifier 11519218.
- Licence recorded as CC BY.
- The licence permits redistribution, so the abstract is shown in full and the full text is available from the publisher.
Abstract
To characterize the 3D geometry of the distal tibia resection area from healthy individuals using CT-based digital implantation for proper preoperative sizing of TAA tibia component placement. Standardized CT images of healthy ankle joints serving as intra-individual references for treatment of contralateral injuries were identified. The tibial cross section dedicated to virtually host the tibial component was digitally prepared, and the size of the virtual contact surface was calculated. Finally, out of five prototypes the one fitting best in terms of size and alignment was identified. CT scans taken from 319 subjects were used for the virtual implantation procedure. Body height and size of the distal tibia contact area correlated (r = 0.49 and 0.42 in females and males, each p < 0.001). Prosthesis sizes 2 and 3 fit well for the vast majority of patients, while the smallest and largest sizes are rarely required. Digital implantation of the tibial component should be considered a valuable tool for preoperative planning as well as for the development of new implant types.
Medical subject headings
- Tomography, X-Ray Computed
- Tibia
- Arthroplasty, Replacement, Ankle
Anatomy
- ankle
- tibia