Clinical evaluation of a multiplex PCR-based test for joint infection: a prospective diagnostic accuracy study of forty-nine patients.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 39358640.
- Also identified by DOI 10.1007/s00590-024-04114-2 and PMC identifier 11519097.
- Licence recorded as CC BY.
- The licence permits redistribution, so the abstract is shown in full and the full text is available from the publisher.
Abstract
The purpose of this study was to evaluate the diagnostic accuracy (sensitivity, specificity, positive predictive value (PPV) and negative predictive value (NPV)) of the PCR-based BioFire® Joint Infection Panel (BJI Panel) against microbiological culture growth for patients suspected of having a native or prosthetic joint infection. Synovial fluid and tissue biopsies were prospectively collected from patients from June 2022 to June 2023. The results of the BJI Panel were compared with those of culture growth. 51 samples were included. Including all pathogens, the sensitivity was 69%, the specificity 89%, the PPV 73% and the NPV 86%. Including only pathogens in the BJI Panel, the sensitivity was 100%, the specificity 90%, the PPV 73% and the NPV 100%. The BJI Panel has a high accuracy for detecting the pathogens in its panel, but the absence of important common pathogens from the panel reduces its sensitivity and NPV. With a short turnaround time and precise pathogen detection, the BJI Panel has the potential to add value as a complementary diagnostic method.
Medical subject headings
- Prosthesis-Related Infections
- Sensitivity and Specificity
- Synovial Fluid
- Multiplex Polymerase Chain Reaction