Performance of Lung Nodule Computer-Aided Detection Systems on Standard-Dose and Low-Dose Pediatric CT Scans: An Intraindividual Comparison.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 39382534.
- Also identified by DOI 10.2214/AJR.24.31972.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
<b>BACKGROUND.</b> When lung nodule computer-aided detection (CAD) systems are applied for pediatric CT, performance may be degraded on low-dose scans due to increased image noise. <b>OBJECTIVE.</b> The purpose of this study was to conduct an intraindividual comparison of the performance for lung nodule detection of two CAD systems trained using adult data between low-dose and standard-dose pediatric chest CT scans. <b>METHODS.</b> This retrospective study included 73 patients (32 female participants, 41 male participants; mean age, 14.7 years; age range, 4-20 years) who underwent both clinical standard-dose and investigational low-dose chest CT examinations during the same encounter from November 30, 2018, to August 31, 2020, as part of an earlier prospective study. Fellowship-trained pediatric radiologists annotated lung nodules to serve as the reference standard. Both CT scans were processed using two publicly available lung nodule CAD systems previously trained using adult data: FlyerScan (github.com/rhardie1/FlyerScanCT) and Medical Open Network for Artificial Intelligence (MONAI; github.com/Project-MONAI/model-zoo/releases). The sensitivities of the two CAD systems for nodules measuring 3-30 mm (<i>n</i> = 247) were calculated when operating at a fixed frequency of two false-positives per scan. <b>RESULTS.</b> FlyerScan exhibited detection sensitivities of 76.9% (190/247; 95% CI, 73.3-80.8%) on standard-dose scans and 66.8% (165/247; 95% CI, 62.6-71.5%) on low-dose scans. MONAI exhibited detection sensitivities of 67.6% (167/247; 95% CI, 61.5-72.1%) on standard-dose scans and 62.3% (154/247; 95% CI, 56.1-66.5%) on low-dose scans. The number of detected nodules for standard-dose versus low-dose scans for 3-mm nodules was 33 versus 24 (FlyerScan) and 16 versus 13 (MONAI), 4-mm nodules was 46 versus 42 (FlyerScan) and 39 versus 30 (MONAI), 5-mm nodules was 38 versus 33 (FlyerScan) and 32 versus 31 (MONAI), and 6-mm nodules was 27 versus 20 (FlyerScan) and 24 versus 24 (MONAI). For nodules measuring 7 mm or larger, detection did not show a consistent pattern between standard-dose and low-dose scans for either system. <b>CONCLUSION.</b> Two lung nodule CAD systems showed decreased sensitivity on low-dose versus standard-dose pediatric CT scans obtained in the same patients. The reduced detection at low dose was overall more pronounced for nodules measuring less than 5 mm. <b>CLINICAL IMPACT.</b> Caution is needed when using low-dose CT protocols in combination with CAD systems to help detect small lung nodules in pediatric patients.
Medical subject headings
- Tomography, X-Ray Computed
- Radiographic Image Interpretation, Computer-Assisted
- Solitary Pulmonary Nodule
- Lung Neoplasms