In vivo imaging of the human brain with the Iseult 11.7-T MRI scanner.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 39420141.
- Also identified by DOI 10.1038/s41592-024-02472-7 and PMC identifier 11541209.
- Licence recorded as CC BY.
- The licence permits redistribution, so the abstract is shown in full and the full text is available from the publisher.
Abstract
The understanding of the human brain is one of the main scientific challenges of the twenty-first century. In the early 2000s, the French Atomic Energy Commission launched a program to conceive and build a human magnetic resonance imaging scanner operating at 11.7 T. We have now acquired human brain images in vivo at such a magnetic field. We deployed parallel transmission tools to mitigate the radiofrequency field inhomogeneity problem and tame the specific absorption rate. The safety of human imaging at such high field strength was demonstrated using physiological, vestibular, behavioral and genotoxicity measurements on the imaged volunteers. Our technology yields T<sub>2</sub> and T<sub>2</sub><sup>*</sup>-weighted images reaching mesoscale resolutions within short acquisition times and with a high signal and contrast-to-noise ratio.
Medical subject headings
- Magnetic Resonance Imaging
- Brain