Removal of errant coils using the Tigertriever-13.
Where this comes from
- Record sourced from PubMed, PMID 39433324.
- Also identified by DOI 10.1136/jnis-2024-021923 and PMC identifier 12163978.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Coil migration poses a significant intraprocedural risk during coil embolization in interventional neuroradiology procedures. In this technical video we describe the technique of removal of errand coils using the low profile Tigertriever-13 device.1 Traditional methods of errant coil retrieval rely on the use of stent retrievers;2 however, these devices are usually too large to fit through smaller inner diameter 0.013-0.0165 inch coiling catheters, thus requiring upsizing materials in emergency situations.3 We demonstrate the technique of coil retrieval using this low-profile device ex-vivo in silicone flow models, then its application in vivo as a 'bail-out' technique during acute embolization of a ruptured peri-callosal aneurysm which required retrieval of errant coils (video 1). neurintsurg;16/12/1212/V1F1V1Video 1Errant coil retrieval using the Tigertriever device.